Installing the TMD
2-12
Installing the TMD
Thermo Fisher Scientific
REC-4105 Rev L
3. Route incoming power wiring through a conduit hole on the bottom
right of the enclosure
Use 14 AWG stranded copper wire.
Maintain a gap of at least ½ in. from low voltage wires.
Leave enough loose wiring to allow the board to be moved without
having to disconnect wires. (Generally, 8 in. is sufficient.)
4. Connect the GROUND wire to the safety ground terminal located on
the right inside of the chassis.
5. Wire the HOT input power to the TB1 terminal labeled “HOT.”
6. Wire the NEUTRAL input power to the TB1 terminal labeled
“NEUT.”
7. Replace the power cover.
Note
: Make sure the chassis cover is closed before applying power.
This section describes the switches that can be set in the field to adapt
the TMD to your environment. SW1 DIP Switches are located on the
display board, which is on the back of the control unit enclosure door.
SW2 – SW6 are located on the CPU board.
SW1 DIP Switches (on the display board) are used to set the detector
hardware configuration. Switches for enabling password protection
software, speed sensor input, SYNC input, belt direction, measure units,
and cold start are set at the hardware level because these parameters are
generally set once and not changed.
The settings of these switches can be viewed on the TMD front panel
after the system is operational. Refer to Switch Setting.
Table 2-1 describes the functions of the DIP Switch settings on SW1.
Configuring the
TMD
DIP Switch Settings
Summary of Contents for Oretronic III
Page 4: ......
Page 8: ......
Page 14: ......
Page 68: ......
Page 100: ......
Page 114: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......