C-2
Thermo Fisher Scientific
REC-4105 Rev L
Note: Do not tighten the bolts securely at this time.
See Appendix D for the engineering drawing related to electrical
installation.
1. At the TMD control panel, from the RUN menu, scroll to the
CALIBRATION menu press the value up arrow until the display
toggles to Y
2. Press
ENTER
3. Scroll through the CALIBRATION menu until EST CAL displays.
4. Press
ENTER
EST takes three readings; these reading are at an interval that is
dependent on your belt speed. The system then takes an average of
the three readings and displays it on the display screen.
5. The reading should display between 250 and 500. If your reading is
below or above those values, you have to gently move EST
horizontally to reposition it.
6. Press the ENTER key and run the calibration procedure again. Keep
moving the EST and running the calibration until your reading
displays in the desired 250-500 signal range.
7. After the value displays between 250 and 500, securely tighten the
nuts on the bolts that run through the cross support beam and the
EST plate.
8. Run the calibration again to ensure the EST plate did not move
while tightening the nuts. If the reading is not between 250 and 500,
loosen the nuts and start the calibration again beginning with Step 5.
EST can be set to run at specific intervals. The interval range is between
1 and 24 hours. Use the following procedures to set up EST interval for
automatic operation:
1. At the TMD control panel, from the RUN menu, scroll to the
SETUP menu press the value up arrow until the display toggles to
Y.
Electrical
Installation
Calibration
and Setup
Interval Setup
Summary of Contents for Oretronic III
Page 4: ......
Page 8: ......
Page 14: ......
Page 68: ......
Page 100: ......
Page 114: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......