137
Power Cords
contact us.
Power Line Conditioning Accessories
A UPS reduces the probability of a system shutdown if power is lost elsewhere in the building.
Power line conditioners (which ensure that your service is free from sags, surges or other line
disturbances) also are available in the U.S.A. from us for 120 volt operation. Line conditioners
for 220 volt operation can be purchased locally. Contact technical support for information
about power conditioners and UPS.
Electrical Service Specifications
The following table lists the specifications for electrical service. Contact our service
representative in your area if you have questions about the requirements.
Power Consumption
Generally, 50% more power should be available than the entire system (including accessories)
typically uses. Maximum power consumption and heat dissipation specifications for the
spectrometer and accessories are shown below. The values are approximate.
Requirements
Specifications
Input current
5.0 A (max.)
Input voltage
100-240 VAC
Line frequency
50-60 Hz
Line disturbances
Sags, surges or other line disturbances must not
exceed 10% of input voltage (even for a half cycle).
Noise
< 2 V (common mode)
< 20 V (normal mode)
Item
Power Consumption
Max. Heat Dissipation
instrument
60 W
205 Btu/hr