Maintenance
Instrument Diagnostics
Thermo Scientific
NanoDrop One
C
with NanoDropQC Software User Guide
125
Before you begin
First make sure the pedestals are properly conditioned. To test pedestal conditioning, clean the
pedestals with a new laboratory wipe, then pipette 1 µL DI H
2
O onto the lower pedestal. The
droplet should “bead up” as shown below. If it does not,
To run performance verification
1. From the instrument home screen, tap
(Diagnostics) and then tap
Performance Verification
. If you are using the PC Control software, from the Home
screen, select
Performance
.
A message asks for target absorbance values.
Note
The PV-1 solution comes in a single-use ampoule. Before you open the ampoule,
shake it vigorously and then allow the liquid to collect in the bottom portion of the
ampoule. After the ampoule is opened, its contents must be used within one hour. Pipette
directly from the ampoule; do not transfer the solution.
Droplet “beads up” on
properly conditioned pedestal