with NanoDropQC Software User Guide
57
Operating Specifications
The instrument operates reliably when the room environment meets these specifications:
• operating temperatures: 5 °C - 35 °C (41 °F - 95 °F)
• relative humidity (non-condensing): 20-80%
Locate the instrument away from air vents and exhaust fans to minimize evaporation.
After the instrument is installed, you can leave it turned on.
Note
If operating the instrument at the low end of the recommended humidity range, use
adequate sample volume to avoid evaporation.