5
Maintenance
Instrument Diagnostics
128
NanoDrop One
C
with NanoDropQC Software User Guide
Thermo Scientific
Swipe left again to see additional measurement details, along with the overall test result.
After the tenth measurement, a message indicates whether the instrument passed or failed
performance verification:
8. If the instrument failed, immediately repeat step 6 using ten 2 µL aliquots of the PV-1
solution.
9. When finished, tap
End Experiment
and
clean the pedestals with 3–5 µL DI H
O
.
Performance test result