Maintenance
Decontaminating the Instrument
120
NanoDrop One
C
with NanoDropQC Software User Guide
Thermo Scientific
To decontaminate the pedestals
To decontaminate the instrument surfaces
1. Lift the instrument arm and clean the upper
and lower pedestal with a new laboratory wipe.
2. Pipette 2–3 µL diluted bleach solution (see
) onto the lower pedestal.
3. Lower the arm and wait 2–3 minutes.
4. Lift the arm and clean both pedestals with a
new wipe.
5. Pipette 3–5 µL DI H
2
O onto the lower
pedestal.
6. Lower the arm and wait 2–3 minutes.
7. Lift the arm and clean both pedestals with a
new wipe.
1. Dampen a clean, soft cloth or laboratory wipe
with the diluted bleach solution (see
) and use it to gently wipe the outside
surfaces of the instrument.
2. Use a clean cloth or wipe dampened with
DI H
2
O to remove the bleach solution.