5
Maintenance
Instrument Diagnostics
Thermo Scientific
NanoDrop One
C
with NanoDropQC Software User Guide
123
To run intensity check
1. Lift the instrument arm and clean the upper and lower pedestal with a new laboratory
wipe.
2. Remove any cuvette from the cuvette holder.
3. Lower the arm.
4. From the instrument home screen, tap
(Diagnostics) and then tap
Intensity Check
. If you are using the PC Control software, from the Home screen, select
Intensity
.
5. On the instrument, Tap
Measure
and wait for the measurements to complete.
Here is an example of a typical intensity check result screen.
If you are using the PC Control software, select
Run
.
6. To rerun the intensity check, tap
Measure
.
7. When finished, tap
End Experiment
.
After the test is completed, the results are available from the Data Viewer (see example
below). See
Manage identifiers on the instrument
for details.
Swipe screen left
to
view detailed results