Dionex AS-DV Operator’s Manual
78
Doc. 065259-05 10/12
4. Insert a 2.5-mm Allen wrench into
the screw on top of the sampling
mechanism (see
) and
turn the wrench clockwise. The
needle begins moving up.
5. Continue turning the wrench until
the top of the sampling head is
even with the top of the sampling
mechanism.
6. If you are clearing a needle jam,
remove the obstruction from the
carousel.
Figure 5-11. Manually Raising the Sampling Head
7. Turn the wrench counterclockwise
to begin lowering the needle. Turn
the wrench until the top of the
sampling head is about 6 mm
(1/4 in) below the top of the
sampling mechanism (see
Figure 5-12. Lowering the Sampling Head
Before Turning on the Power
8. Reinstall the sampling head tower cover.
9. Reinstall the front cover and close the top cover.
10. Reconnect the power cord and turn on the power switch.
Sampling
Mechanism
Sampling
Mechanism
Sampling
Mechanism
Sampling
Head