Dionex AS-DV Operator’s Manual
76
Doc. 065259-05 10/12
11. Thread the free end of the tubing up through the needle and
screw the tip into place on the needle.
12. Continue threading the free end of the tubing up through
the sampling head and out the aperture on the front of the
Dionex AS-DV.
13. Slide the needle into place and slide the sampling needle tubing back into the
slot in the sampling head.
14. Align the flat sides of the sampling tip so that
they fit through the opening in the sensor
block.
15. Make sure that the needle is pushed as far as it
will go up into the sampling head and then
tighten the set screw firmly.
NOTE The set screw must be very tight to
secure the sampling needle.
16. Reinstall the sampling head tower cover.
17. Reinstall the front cover and close the top cover.
18. Reconnect the sample out line to the valve.
19. Reconnect the power cord and turn on the power.
Opening in
Sensor Block