Dionex AS-DV Operator’s Manual
106
Doc. 065259-05 10/12
•
Normal Pulse
: In normal pulse
operation, the negative (falling)
edge of the TTL signal is the
active edge and the positive
(rising) edge is ignored.
A pulse width of 50 ms or more is guaranteed to be detected. A pulse
width of 4 ms or less is guaranteed to be ignored. The action for pulse
widths that are greater than 4 ms and less than 50 ms is undefined.
•
Inverted Pulse
: The inverted
pulse mode operates identically
to the normal pulse mode
except that the positive and
negative edges are reversed in
function.
Selecting the Control Mode
1. Start the Chromeleon 7 Instrument Configuration Manager or the
Chromeleon 6.8
Server Configuration program:
•
For Chromeleon 7, click
Start
on the Windows taskbar and click
All Programs > Chromeleon 7 > Instrument Configuration
Manager
.
•
For Chromeleon 6.8, click
Start
on the Windows taskbar and
click
All Programs > Chromeleon > Server Configuration
.
2. Under the instrument or timebase, double-click the
Dionex AS-DV
icon.
The Dionex AS-DV Properties dialog box appears.
3. Click the
TTL Inputs
tab and then press the
F2
key.