5 • Service
Doc. 065259-05 10/12
73
10. Unscrew the thumbscrew
on the carousel, and lift the
carousel out of the
autosampler.
Figure 5-6. Removing the Carousel
11. Clear any blockages and clean the basin.
12. Align the holes on the bottom of the carousel
with the pins on the top of the carousel shaft
and reinstall the carousel. Replace the
thumbscrew and hand tighten.
13. Slide the needle tubing back into the slot in the
sampling head.
14. Reinstall the sensor block.
15. Align the flat sides of the sampling tip so that
they fit through the opening in the sensor
block.
16. Make sure that the needle is pushed as far as it
will go up into the sampling head and then
tighten the set screw firmly.
NOTE The set screw must be very tight to
secure the sampling needle.
17. Reinstall the sampling head tower cover.
18. Reinstall the front cover and close the top cover.
19. Reconnect the power cord and turn on the power.
Alignment Pins on Carousel
Opening in
Sensor Block