Thermo Scientific APEX User’s Guide
Page 209
Why Locking the AuditCheck Values Is Useful
Locking the AuditCheck values for phase angle and peak signal is useful, because it
prevents the less-accurate auto-calibration values over-writing the more-accurate
AuditCheck values. Auto-calibration values are inherently more variable, because the
response of the search head to the passage of a contaminant is dependent on position—
for the simple reason that the edge of the head is significantly more responsive than the
center. As a result, the AuditCheck values are significantly more accurate than the auto-
calibration values for the following reasons. First, the AuditCheck sample always passes
through the head in exactly the same position; and second, the AuditCheck sample
passes through the edge of the search head, where its response to metallic
contaminants is greatest.
Locking the AuditCheck Values
1)
To lock the AuditCheck values for phase angle and peak signal, highlight the padlock
function.
2)
Press the Go button and the X changes to a check mark.
3)
Press the Back button to lock these values and exit the function.
Setting Manual-Start, Timing-Limits, and Reject-Device
Parameters
These functions have already been described in detail for the Quality Test function in the
Quality Test section, starting on page 196—so please refer to the appropriate sections in
the Quality Test write-up for detailed instructions, as the Timing-Limits, and Reject
Device Parameters are very similar for AuditCheck and the Quality Test function.
AuditCheck – Timers and Reject
(continued…)
Summary of Contents for APEX 100
Page 1: ...Thermo Scientific APEX Metal Detector User s Guide REC 4248 Rev K Part number 085381 English ...
Page 10: ...Page 10 Thermo Scientific APEX User s Guide ...
Page 78: ...Page 78 Thermo Scientific APEX User s Guide ...
Page 110: ...Page 110 Thermo Scientific APEX User s Guide ...
Page 258: ...Page 258 Thermo Scientific APEX User s Guide ...
Page 270: ...Page 270 Thermo Scientific APEX User s Guide ...
Page 282: ...Page 282 Thermo Scientific APEX User s Guide ...
Page 310: ...Page 310 Thermo Scientific APEX User s Guide ...
Page 319: ...Thermo Scientific APEX User s Guide Page 319 Appendix A Modbus Communications Card ...
Page 320: ...Page 320 Thermo Scientific APEX User s Guide ...
Page 324: ...Page 324 Thermo Scientific APEX User s Guide ...
Page 340: ...Page 340 Thermo Scientific APEX User s Guide ...
Page 367: ...Thermo Scientific APEX User s Guide Page 367 Appendix B IntelliTrack XR ...
Page 368: ...Page 368 Thermo Scientific APEX User s Guide ...
Page 402: ...Page 402 Thermo Scientific APEX User s Guide ...
Page 404: ...This page intentionally left blank ...
Page 414: ...This page intentionally left blank ...
Page 430: ......
Page 431: ......
Page 432: ......
Page 433: ......
Page 434: ......
Page 435: ......
Page 436: ......
Page 437: ......
Page 438: ......
Page 439: ......
Page 440: ......
Page 441: ......
Page 442: ......