Operation
Settings
Thermo Scientific
43iQ Instruction Manual
3-73
●
Lamp Voltage:
Displays the current lamp voltage.
●
Lamp Intensity:
Displays the current lamp intensity.
●
Down:
●
Manual Entry:
Shows current value. If pressed, user can manually
set the detector gain value.
●
Auto Cal:
When pressed, the detector’s auto-calibration process is
initiated. Please allow up to 5 minutes for calibration to complete.
User can stop calibration by pressing the Stop button under
Detector Gain.
●
End Cal:
When pressed, the auto calibration is interrupted and the
value does not change.
●
Default Values:
When pressed, the default gain for the lamp
intensity detector is restored.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue To PMT Settings Screen
The PMT Settings screen contains the following information:
●
Across:
●
Description:
Defines the actions the user can do.
●
PMT Supply:
User can manually adjust the PMT supply value, read
target and calibration concentration, start/stop automatic
calibration, read voltage, frequency and zero offset frequency
values, and reset default values.
●
Current PMT:
Displays the current PMT voltage value.
●
Units:
Displays units for each item.
Summary of Contents for 43iQ
Page 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Page 2: ......
Page 8: ......
Page 18: ......
Page 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Page 162: ......
Page 170: ......
Page 180: ......
Page 228: ......
Page 238: ......
Page 262: ......
Page 270: ......
Page 282: ......
Page 283: ......