Servicing
DMC Optical Bench Removal
Thermo Scientific
43iQ Instruction Manual
7-39
Use the following procedure to replace the kicker assembly (
).
Equipment required:
Phillips drive, #1
1.
Disconnect plumbing as shown.
2.
Loosen 4-40 screw. Slide kicker assembly towards opening of keyway.
Equipment Damage
Do not loosen or tighten this side of fitting as the
internal tubing will twist blocking flow. These fittings are covered by
silicone tape as a reminder.
▲
3.
Once kicker assembly clears the keyway, pull outwards towards the
board and up.
Figure 7–44.
Removing and Replacing the Kicker Assembly
Kicker Assembly
Replacement
Summary of Contents for 43iQ