Introduction
Specifications
Thermo Scientific
43iQ Instruction Manual
1-5
lists the specifications for the 43iQ.
Table 1–1.
43iQ Specifications
Range
0
−
10 ppm
0
−
25 mg/m
3
Extended Ranges
0
−
100 ppm
0
−
250 mg/m
3
Zero Noise
1.0 ppb RMS (10 second averaging time)
0.5 ppb RMS (60 second averaging time)
0.25 ppb RMS (300 second averaging time)
Detection Limit
2 ppb (10 second averaging time)
1 ppb (60 second averaging time)
0.25 ppb (300 second averaging time)
Zero Drift
<0.5 ppb (24 hour)
Span Drift
±
0.5% full-scale (24 hour)
Response Time
60 sec (10 second averaging time)
110 sec (60 second averaging time)
320 sec (300 second averaging time)
Linearity
±
1% full-scale
Flow Rate
0.5 lpm (standard)
1.0 lpm (optional)
Interferences (EPA levels)
Less than lower detectable limit except for the following:
NO < 3 ppb, M-Xylene < 1ppb, H
2
O < 3% of reading
Operating Temperature Range 0–
45
°
C
Power Requirements
100–240 VAC 50/60 Hz
275 watts
Physical Dimensions
24 in (D) x 16.75 in (W) x 8.72 in (H) [609 mm (D) 425.45 mm
(W) x 221.48 mm (H)]
Weight
37.1 lbs
Analog I/O
4 Isolated Voltage Inputs 0–10 V
6 Isolated Analog Voltages Outputs, with 4 selectable
ranges
6 Isolated Analog Current Outputs, with 2 selectable ranges
Digital I/O
16 Digital Inputs (TTL)
8 Solenoid Driver Outputs
10 Digital Reed Relay Contact Outputs
Specifications
Summary of Contents for 43iQ
Page 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Page 2: ......
Page 8: ......
Page 18: ......
Page 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Page 162: ......
Page 170: ......
Page 180: ......
Page 228: ......
Page 238: ......
Page 262: ......
Page 270: ......
Page 282: ......
Page 283: ......