Troubleshooting
Troubleshooting Guide
6-6
43iQ Instruction Manual
Thermo Scientific
Problem
Possible Cause
Action
possible.
Noise pick-up by recorder or
data logger
Check analog cable shielding
and replace.
Try to localize source of noise
by comparing analog signal to
data collected through RS-232
or Ethernet.
Poor linearity
Problem with calibrator
Verify accuracy of the
calibration system with an
independent flow meter.
Hydrocarbon kicker
Replace kicker if older than 3
years.
Leak in sample probe line
Check for variable dilution.
Excessive response time
Averaging time is not set
correctly
Check the “Averaging Time” in
the Settings>Measurement
Settings screen.
Instrument is not drawing in
sample at the expected
flow rate
Check sample flow and
pressure readings in the
Settings>Health Check>Status
and Alarms>Flow and Pressure
screen.
Use an independent flow meter
to check flows at the sample
inlet and exhaust bulkheads
(they should match).
Perform a leak test as
described on page
SO
2
is being absorbed and
released by dirt in tubing or
filters of the calibration
system or contamination
inside the instrument
Replace any lines made of vinyl
or plastics with fresh PTFE or
stainless steel.
Replace PTFE filter membranes
that look dirty.
Remove any filters that are not
PTFE membranes.
Contaminated bench
Contact technical support for
additional information and
options.
Analog signal doesn’t
match expected value
Software has not been
configured
Verify that the selected analog
output has been properly
configured to match the data
system.
Analog output goes above
By default, a 5% over and
Summary of Contents for 43iQ
Page 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Page 2: ......
Page 8: ......
Page 18: ......
Page 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Page 162: ......
Page 170: ......
Page 180: ......
Page 228: ......
Page 238: ......
Page 262: ......
Page 270: ......
Page 282: ......
Page 283: ......