Operation
Calibration
Thermo Scientific
43iQ Instruction Manual
3-13
The Calibrate Span Coefficient screen contains the following information:
●
Edit Span Concentration:
User enters the span gas concentration when
in single range mode.
●
Edit High Range Span Concentration:
User enters the high range span
concentration when in dual or auto range mode.
●
Edit Low Range Span Concentration:
User enters the low range span
concentration when in dual or auto range mode.
●
Current High Range Concentration:
Read only. Current high range
concentration reading when in dual or auto range mode.
●
Current Low Range Concentration:
Read only. Current low
concentration reading when in dual or auto range mode.
●
Current Span Coefficient:
Read only. Displays the current instrument
span coefficient.
●
Current High Range Span Coefficient:
Read only. Displays the current
instrument high range span coefficient.
●
Current Low Range Span Coefficient:
Read only. Displays the current
instrument low range span coefficient.
●
Calculated Span Coefficient:
Read only. After the “Edit Span
Concentration” value is entered, the new calculated span coefficient is
displayed.
●
Calculated High Range Span Coefficient:
Read only. After the “Edit
High Range Span Concentration” value is entered, the new calculated
high range span coefficient is displayed.
●
Calculated Low Range Span Coefficient:
Read only. After the “Edit Low
Range Span Concentration” value is entered, the new calculated low
range span coefficient is displayed.
●
Calibrate:
When pressed, updates the coefficient and the concentration
should match the span concentration.
Summary of Contents for 43iQ
Page 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Page 2: ......
Page 8: ......
Page 18: ......
Page 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Page 162: ......
Page 170: ......
Page 180: ......
Page 228: ......
Page 238: ......
Page 262: ......
Page 270: ......
Page 282: ......
Page 283: ......