Operation
Calibration
Thermo Scientific
43iQ Instruction Manual
3-21
●
Adjust High Range Span Coefficient:
User manually adjusts high range
span coefficient when in dual or auto range mode.
●
Adjusted High Range Concentration:
Read only. Shows adjusted high
range concentration based on adjusted high range span coefficient
when in dual or auto range mode.
●
Adjust Low Range Span Coefficient:
User manually adjusts low range
span coefficient when in dual or auto range mode.
●
Adjusted High Range Concentration:
Read only. Shows adjusted low
range concentration based on adjusted low range span coefficient when
in dual or auto range mode.
●
Calibrate:
Calibrates span coefficient by saving the newly adjusted span
coefficient.