Calibration
Manual Calibration
Thermo Scientific
43iQ Instruction Manual
4-13
Concentration shows what the new SO
2
concentration would be based on
the adjusted span coefficient. Press the Calibrate button to save the
adjusted span coefficient value.
Home Screen>Calibration>Advanced>Manual Calibration>Adjust
Span Coefficient (single range mode)
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust High Range Span Coefficient (dual or auto range
mode)
The Reset Bkg to 0.000 and Span Coef to 1.000 screen allows the user to
reset the calibration configuration values to factory defaults.
Reset Bkg to 0.000
and Span Coef to
1.000