Servicing
I/O Replacement
7-14
43iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the I/O boards.
1.
Turn instrument OFF, unplug power cord, and remove the cover
2.
Swing arm open.
Figure 7–15.
I/O Replacement, Arm
3.
Pull board upwards.
Figure 7–16.
I/O Replacement, Remove Board
I/O Replacement