Operation
Settings
3-76
43iQ Instruction Manual
Thermo Scientific
The Pressure Calibration screen is used to calibrate the pressure sensor to
zero, span, or factory default values.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Pressure Calibration
The Pressure Calibration screen contains the following information:
●
Across:
●
Description:
Lists items in table.
●
Reading:
Displays reading of each pressure sensor.
●
Calibration:
Starts calibration or resets default values.
●
Down:
●
Sensor 1
−
3 Reading:
Under the column labeled Reading, current
reading of each pressure sensor.
●
Atmospheric Sensor 1
−
3:
Under the column labeled reading, the user
enters the current atmospheric pressure in mmHg units. Under the
column labelled Calibration, the user presses the Start button to
calibrate the high point of the sensor.
●
Zero Sensor 1
−
3:
User should put the pressure sensor under a strong
vacuum. Under the column labeled reading, the user enters the
pressure in mmHg. Under the column labeled Calibration, the user
presses the Start button to calibrate the low point of the sensor.
●
Reset all values:
Resets values to default.
Pressure Calibration
Summary of Contents for 43iQ
Page 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Page 2: ......
Page 8: ......
Page 18: ......
Page 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Page 162: ......
Page 170: ......
Page 180: ......
Page 228: ......
Page 238: ......
Page 262: ......
Page 270: ......
Page 282: ......
Page 283: ......