64
3-4-2. Select Horizontal Scale
Select
Horizontal Scale
To select the timebase (time/div), turn
the horizontal Scale knob; left (slow)
or right (fast).
SCALE
Range
5ns/div ~ 100s/div, 1-2-5 increment
The timebase indicator updates as the horizontal
scale is adjusted.
Horizontal
position
Timebase
Run Mode
In Run mode, the memory bar and waveform size
keep their proportion. When the timebase
becomes slower, roll mode is activated (if the
trigger is set to Auto).
Stop Mode
In Stop mode, the waveform size changes
according to the scale.
3-4-3. Select reference point of the horizontal expansion
Background
Tthe ability to change the center point of the
horizontal expansion.
Panel Operation 1. press the Acquire key
Acquire
2. Press the Expand key and choose
By Center or By Trigger Pos.