94
CAN
Start of Frame, Type of Frame,
Identifier, Data, Id & Data, End of
Frame, Missing Ack, Bit Stuffing Err
LIN
Sync, Identifier, Data, Id & Data,
Wakeup Frame, Sleep Frame, Error
Threshold
(Pulse Runt)
Sets the upper threshold limit.
Sets the lower threshold limit.
Threshold
(Rise & Fall)
Sets the High threshold.
Sets the Low threshold.
Trigger When
(Timeout)
Stays High Triggers when the input signal stays
high for a designated amount of time.
Stays Low Triggers when the input signal stays
low for a designated amount of time.
Either
Triggers when the input signal stays
high or low for a designated amount of
time.
Timer
(Timeout)
4nS~10.0S Sets the amount of time that a signal
must stay high or low for the timeout
trigger.
3-7-3. Setup Holdoff Level
Background
The holdoff function defines the waiting period
before the DCS-1000B starts triggering again after
a trigger point. The holdoff function ensures a
stable display if there are a number of points in a
periodic waveform that can be triggered. Holdoff
applies to all the triggering types.
Wait for
trigger
Holdoff time
Trigger point
Next trigger point
Panel Operation 1. Press the trigger
Menu
key.
Menu
2. To set the Holdoff time, press the
Holdoff
(or
Mode/Holdoff
) menu
button on the bottom bezel.
3. Use the side menu to set the
Holdoff time.
Range
4ns~10s