99
Odd Field NTSC: 1 ~ 263
PAL/SECAM: 1 ~ 313
EDTV: 1~525(480P), 1~625(576P)
HDTV: 1~750(720P), 1~563(1080i),
1~1125(1080P)
Even
Field
NTSC: 1 ~ 262
PAL/SECAM: 1 ~ 312
HDTV: 1~562(1080i)
All Fields Triggers on all fields.
All Lines Triggers on all lines.
8. Press
Polarity
to toggle the polarity
type.
Range
positive, negative
3-7-9. Pulse Runt trigger
Panel Operation 1. Press the trigger
Menu
key.
Menu
2. Press the
Type
key from the lower
bezel menu.
3. Select
Others
→
Pulse
Runt
from the side menu.
The Pulse and Runt
indicator appears at the
bottom of the display.
From left: polarity, source, high/low threshold,
threshold level, coupling
4. Press
Source
from the lower menu.
Use the side menu to select a source.
Range
Channel 1 ~ 4(Alternate On/Off)
5. Press
Polarity
to toggle the polarity.
Range
Rising edge, falling edge, either.
6. Press
When
from the lower menu.
Then use the side menu to select the condition
and width.
Condition > , < , =
, ≠