78
Polarity Normal (High = 0), Inverted (High = 1)
Configuration
The Configure key sets the baud rate, number of
data bits and parity.
1. Press
Configure
from the bottom
menu.
2. From the side menu select the
Baud rate
,
Data
bits
,
Parity
,
Packets
and
End of Packet bits
.
Baud
Rate
50, 75, 110, 134, 150, 300, 600,
1200, 1800, 2000, 2400, 3600,
4800, 7200, 9600, 14400, 15200,
19200, 28800, 31250, 38400,
56000, 57600, 76800, 115200,
128000, 230400, 460800, 921600,
1382400, 1843200, 2764800
Data Bits 8 (fixed)
Parity
Odd, Even, None
Packets
On, Off
End of
Packet
(Hex)
00(NUL), 0A(LF), 0D(CR), 20(SP),
FF
3-6-2-3. I
2
C Serial Bus Interface
The I
2
C bus is a 2 wire interface with a serial data line (SDA) and
serial clock line (SCLK). The I
2
C protocol supports 7 or 10 bit
addressing and multiple masters. The scope will trigger on any of
the following conditions: a start/stop condition, a restart, a missing
acknowledge message, Address, Data or Address&Data frames.
The I
2
C trigger can be configured for 7 or 10 bit addressing with the
option to ignore the R/W bit as well as triggering on a data value or
a specific address and direction (read or write or both).
Panel operation 1. Connect each of the bus signals (
SCLK
,
SDA
)
to one of the oscilloscope channels. Connect
the ground potential
to one of the probes’
ground clip.
GND
SCLK
SDA
X10
X1
X10
X1