99
Alignment
The alignment of the KnieAgil® P41
should be based on the TKA line. The load
line of the socket should run along the
front edge of the superior anterior axis.
The AP position of the adapter is adju
stable, which also allows for adjustments
of the load line.
Fastening torque of screw
connection between adapter and
knee joint must be 20 Nm!
Static Fitting
Upon delivery, every joint is set to
a basic setting.
Ask the patient to stand up. Remain close,
so you can support the patient if necessary.
Individually adjust all components until
the patient is able to stand securely.
Optional adjustments are described in
the section “Dynamic Fitting”.
EN
FR
ES
IT
DD
DE
KnieAgil® Joints
User Manual P41
Scope of Delivery
REF
Product
KA.P41
KnieAgil® P 41
90.395.093.01
User Manual
Accessories
REF
Product
Required Accessories
KA.KPA-33A
Pyramid Adapter
KA.KPA-
SET-03-34
Lamination Anchor
KA.KPA-21
Threaded Connector
36 mm
Optional Accessoires
21.989.002.00
Cosmetic Foam Cover
Application/Adjustment
We recommend executing
the following steps of the
application process in the
order suggested below.