111
EN
FR
ES
IT
DD
DE
KnieAgil® Joints
User Manual PF50
Scope of Delivery
REF
Product
KA.PF50
KnieAgil® PF 50
90.395.093.01
User Manual
Accessories
REF
Product
Optional Accessoires
21.989.003.00
Cosmetic Foam Cover
KA.EA075
Spring for
KA.PF50, soft
KA.EA052
Spring for KA.PF50,
medium (same as
preinstalled spring
on joint on delivery)
KA.EA053
Spring for
KA.PF50, hard
KA.AG115
Bumper for Stance
Flexion Unit (2 pcs)
Application/Adjustment
We recommend executing
the following steps of the
application process in the
order suggested below.
Alignment
The alignment of the KnieAgil® PF50
should be based on the TKA line. The load
line of the socket should run through the
superior anterior axis.
Static Fitting
Upon delivery, every joint is set to a
basic setting.
Ask the patient to stand up. Remain close,
so you can support the patient if necessary.
Individually adjust all components until
the patient is able to stand securely.
Optional adjustments are described in
the section “Dynamic Fitting”.