Maintenance
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
120 of 158
Maintenance
Scheduled Maintenance Overview
After Each Test
Clean up any spills or dissolution media wherever it was spilled
on the system. Wipe down the system with a clean damp cloth
when finished.
Rinse baskets, basket shafts, paddles, and/or vessels with DI
or better quality water. Avoid harsh detergents which may
damage the materials.
Weekly
Inspect bath water for algae or other contaminants. Drain and
replace if necessary.
Monthly
Replace bath water, regardless of condition, to prevent algae
growth.
Every 6 Months
Check the temperature and motor speed calibration.
Inspect the spindles for smooth rotation.
Inspect the o-rings to ensure that the spindle shafts are held
securely.
Annually
Inspect the drive belts and idler pulleys.
Every Two Years
Replace drive belts.
Summary of Contents for VISION G2 CLASSIC 6
Page 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Page 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Page 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Page 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Page 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Page 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Page 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Page 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...