Troubleshooting
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
127 of 158
Heater Troubleshooting
Problem
Possible Causes
Recommended Action
Tester doesn't
communicate with heater
Heater not turned on.
Turn on heater.
Heater not connected.
Connect heater to tester using
RS-232 cable to Temp Control
ports.
Board failure.
Contact THR Technical
Support.
Main Check Error when
tester heats up beyond a
certain temperature
Component failure.
Contact THR Technical
Support.
Tester heats up to
different temperature than
set point
Set point has changed due to
test starting.
Confirm set point on main
screen.
Offset for heater probe is
incorrect.
Correct heater probe offset on
probe calibration screen.
Heater probe has failed
Replace heater probe.
External temperature probe
offset is incorrect.
Correct external probe offset
on probe calibration screen.
External temperature probe
has failed.
Replace external temperature
probe.
When measuring
temperature, Probe Error
or --.-- is displayed
Probe is disconnected.
Reconnect temperature probe.
Probe is damaged.
Contact THR Technical
Support.
Summary of Contents for VISION G2 CLASSIC 6
Page 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Page 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Page 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Page 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Page 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Page 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Page 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Page 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...