Electrical Troubleshooting
Problem
Possible Causes
Recommended Action
Display is dark
Tester is not turned on.
Use power switch to turn
tester on.
Tester is not plugged in.
Ensure tester is plugged into
an outlet that is delivering
power.
USB functions not working USB Drive is not formatted.
The Teledyne Hanson USB
drive has already been
formatted. If using another
USB drive, format using a PC.
USB drive is damaged.
Use another USB drive.
Monument board is
damaged.
Replace monument board.
System does not make
noise
Speaker volume set to 0.
Use Display screen to raise
speaker volume.
Display board is damaged.
Replace display board.
Screen is illuminated but
system not responsive
System has frozen.
Note last action, reboot
system. Report issue to THR
Technical Support.
Touchscreen requires
calibration.
If possible, recalibrate through
display screen. If display
screen cannot be reached,
reboot system and touch the
screen during the logo splash
screen.
Summary of Contents for VISION G2 CLASSIC 6
Page 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Page 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Page 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Page 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Page 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Page 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Page 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Page 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...