NOTE: If a Quick Start speed is not set, the Elapsed Time field
will not respond when touched.
To adjust the temperature, touch the current temperature on the
home screen and a numeric keypad will appear. If there is a
problem with the external temperature probe, “Probe Error” will
be displayed.
Enter the desired temperature and touch Ok. The pump will start
and the temperature should begin to rise. There are preset
buttons for 32 °C and 37 °C. The PUMP button starts the pump
if it is off, or stops it if it is running. The temperature will
automatically be set to 0 if the pump is turned off. The OFF
button turns off the pump and sets the temperature to 0.
NOTE: As a safety feature, manual control of the pump is only
active when the set temperature is 0.
Summary of Contents for VISION G2 CLASSIC 6
Page 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Page 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Page 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Page 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Page 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Page 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Page 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Page 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...