Operation
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
77 of 158
The protocols are listed alphabetically by protocol name in
groups of 5 over 5 pages within the interface. The options
available for protocols are shown on the buttons on the right
side of the screen when all protocols are displayed:
Checksum button
This button generates a unique checksum for the protocols.
This can be used to ensure a restored set of protocols is
identical to the original set. When touched, the Protocols check
pop-up is displayed, which first informs the user the value is
being calculated, then displays the calculated value.
For example, an operator touches the Checksum button and
obtains a value of 0EA4. The operator then changes a protocol
label by 1 character. When the operator touches the Checksum
button again, it will generate a different value.
This feature can be used to ensure that backups and
restorations of protocols are identical to the originals. The
feature can also be used by managers to ensure that protocols
have not been changed without authorization.
Summary of Contents for VISION G2 CLASSIC 6
Page 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Page 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Page 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Page 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Page 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Page 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Page 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Page 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...