Installation
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
31 of 158
Priming the Waterbath
Method 1
WARNING:
Do not set a temperature during this procedure.
Doing so may result in damage to the Vision heater.
1.
Fill the waterbath to the line on the sticker located at
the center front of the bath.
Fig. C-8 Fill waterbath to appropriate line on sticker
Go to the Temperature Setpoint screen and touch the PUMP
button.
WARNING:
Do not set a temperature during this procedure.
Doing so may result in damage to the Vision heater.
Completely fill the syringe with bath water.