ADFM Commands
EZ - Sensor Source:
Purpose:
The EZ-command selects the source of environmental sensor
data.
Format: EZcdhprst
Default EZ1011111
Range:
Firmware switches (see description)
Description
Setting the EZ-command firmware switches tells the ADFM
to use data from a manual setting or from an associated sen-
sor. When a switch value is non-zero, the ADFM overrides
the manual E-command setting and uses data from the appro-
priate sensor. If no sensor is available, the ADFM defaults to
the manual E-command setting. The following table shows
how to interpret the sensor source switch settings.
Table C-5:
Sensor Source Switch Settings
Field
Value = 0
Value = 1
Value = 2
c
Manual EC
Calculate using
ED, ES, and ET
N/A
d
Vertical beam only
Vertical beam
and secondary
depth sensor
N/A
h
Manual EH
Heading
N/A
p
Manual EP
Pitch (Tilt 1)
N/A
r
Manual ER
Roll (Tilt 2)
N/A
s Manual
ES
N/A
N/A
t Manual
ET
Transducer
Sampled
N/A
Example
EZ1111101 means calculate speed of sound from readings,
use secondary depth sensor, transducer heading, internal tilt
sensors, and transducer temperature.
ADFM Technical Manual (January 2000)
page C-17
Summary of Contents for ADFM Analog Output Module
Page 2: ......
Page 6: ...Table of Contents NOTES iv MGD Technologies Inc ...
Page 13: ...Table of Contents NOTES ADFM Technical Manual January 2000 xi ...
Page 14: ...Table of Contents xii MGD Technologies Inc ...
Page 54: ...Chapter 5 NOTES page 5 10 MGD Technologies Inc ...
Page 132: ...Appendix C NOTES page C 52 MGD Technologies Inc ...
Page 152: ...Appendix E NOTES page E 14 MGD Technologies Inc ...
Page 154: ......
Page 155: ......
Page 156: ......