ADFM Commands
Timing Card RAM (PI ): PASS
Demodulator RAM (PI ): PASS
Checksum Code/Tables (PT8): PASS
Receive Test (PT3): PASS
Transmit Test (PT4): PASS
Electronics Wrap Test (PT5): PASS
LPF Bandwidth Test (PT6): PASS
Clock Interrupt (PI ): PASS
Error Log:
Power Loss
Auto Restart Occurred
Transducer Communications Error
Self Tests Complete
PD - Data Stream Select
Purpose:
Selects the type of ensemble output data structure.
Format: PDnn
Range
n = 0 to 3 (see description)
Default PD00
Description: PD selects either the normal output data structure, a special
application data structure, or a fixed data set for transmis-
sion/display as the data ensemble.
Table C-6:
Data Stream Selections
Format Description
PD0
Sends The real water-current data set
PD1
Sends an RDI-defined data set that always uses the same data (except
for parts of the leader data). This data set is useful during user-software
development.
PD2 Not
used.
PD3
Sends reduced discharge output data structure.
PF - Display Fault Log
Purpose:
The PF-command displays the fault log.
Format: PF
Description: Displaying the fault log will list why a built-in test failed.
This may aid in troubleshooting.
Example:
>pf
Power Loss
Auto Restart Occurred
Transducer Communications Error
ADFM Technical Manual (January 2000)
page C-23
Summary of Contents for ADFM Analog Output Module
Page 2: ......
Page 6: ...Table of Contents NOTES iv MGD Technologies Inc ...
Page 13: ...Table of Contents NOTES ADFM Technical Manual January 2000 xi ...
Page 14: ...Table of Contents xii MGD Technologies Inc ...
Page 54: ...Chapter 5 NOTES page 5 10 MGD Technologies Inc ...
Page 132: ...Appendix C NOTES page C 52 MGD Technologies Inc ...
Page 152: ...Appendix E NOTES page E 14 MGD Technologies Inc ...
Page 154: ......
Page 155: ......
Page 156: ......