Appendix C
EE - Secondary Depth Max. Probable Error
Purpose:
Sets the maximum probable error of the secondary depth sen-
sor.
Format: EE
nnnn
Range:
nnnn
= 0 to 1000 (o to 100%)
Default: EE0020
(2%)
Description: The EE command allows the user to set the maximum prob-
able error (MPE) of the secondary depth sensor. If the depth
as determined by the vertical beam is within the MPE of the
secondary sensor, the vertical beam depth is used. Otherwise,
the secondary depth sensor depth is used. The secondary
depth sensor must be selected with EZ=x1xxxxx.
EO - Secondary Depth Zero Offset
Purpose:
Sets the secondary depth zero offset in the ADFM.
Format: EO
nnnn
Range:
8 to 9999 mm
Default: EO0008
Description: The secondary depth zero offset determines the distance above
the bottom of the channel at which the secondary depth sensor
is placed. The default setting of 8 mm is half the diameter of
the recommended secondary sensor.
ER - Secondary Depth Span
Purpose:
Sets the operating range of the secondary depth sensor.
Format: ER
nnnnn
Range:
0 to 30000 mm
Default: ER00000
Description: Sets the full-scale depth of the secondary depth sensor in mil-
limeters. If this value is unchanged (i.e. left at zero), all
depths will be zero + zero offset (EO-command).
page C-14
MGD Technologies, Inc.
Summary of Contents for ADFM Analog Output Module
Page 2: ......
Page 6: ...Table of Contents NOTES iv MGD Technologies Inc ...
Page 13: ...Table of Contents NOTES ADFM Technical Manual January 2000 xi ...
Page 14: ...Table of Contents xii MGD Technologies Inc ...
Page 54: ...Chapter 5 NOTES page 5 10 MGD Technologies Inc ...
Page 132: ...Appendix C NOTES page C 52 MGD Technologies Inc ...
Page 152: ...Appendix E NOTES page E 14 MGD Technologies Inc ...
Page 154: ......
Page 155: ......
Page 156: ......