ADFM Commands
C-2.5 Discharge
Commands
The ADFM uses the following commands to calculate the discharge.
QB - Bed Level
Purpose:
Sets the bed level in the pipe or channel.
Format: QB
nnnn
Range:
0 to 2000 mm
Default: QB0000
Description: Use this command if the bottom of the pipe or channel has
filled in with silt. Water flow under the transducer is not pos-
sible and is not calculated for the total discharge.
QC - Channel Type
Purpose:
Sets the pipe or channel type between rectangular or circular.
Format: QC
nnn
Range:
nnn
= 0, 1 (0 = rectangle, 1 = circle)
Default: QC000
Description: Use this command to set the pipe or channel type between rec-
tangular or circular.
QD - Algorithm Select
Purpose:
This command turns on or off various methods used to calcu-
late the flow rate.
Format: QD
abc def ghi
Range:
Firmware switches (see description)
Default: QD100
011
110
Description: See table below for descriptions.
Switch Description
a = Qmain
Do other selected algorithms (order of priority e, f, h, and g) and output a single best answer.
b = not used
c = not used
d = not used
e = IHRW
Data adaptive fit
f = QmodVA
Average Velocity x Average Area with R
1/7
power law applied to velocity near the boundry.
g = QManning
(not implemented)
h = Qva
Average Velocity x Average Area
i = not used
ADFM Technical Manual (January 2000)
page C-33
Summary of Contents for ADFM Analog Output Module
Page 2: ......
Page 6: ...Table of Contents NOTES iv MGD Technologies Inc ...
Page 13: ...Table of Contents NOTES ADFM Technical Manual January 2000 xi ...
Page 14: ...Table of Contents xii MGD Technologies Inc ...
Page 54: ...Chapter 5 NOTES page 5 10 MGD Technologies Inc ...
Page 132: ...Appendix C NOTES page C 52 MGD Technologies Inc ...
Page 152: ...Appendix E NOTES page E 14 MGD Technologies Inc ...
Page 154: ......
Page 155: ......
Page 156: ......