ADFM Commands
QP - Disc Params
Purpose:
Sets the values used in the profiling algorithm and the power
law exponent used in the ModVA flow rate algorithm.
Format: QP
nnnn
,
nnnn
Range:
nnnn
= Delta, N (0 = calculate, not 0 = fixed)
Default: QP0500,0250
Description: Contact MGD Technologies Inc. for further details.
QR - QVD Structure Only Rate
Purpose:
Sets the rate at which full “PD0” zero ensembles are inter-
leaved between “PD3” ensembles.
Format: QR
nnn
Range:
nnn
= 1 to 256 (1 = both always)
Default: QR006
Description: If PD3 is selected, a PD0 ensemble is output for ensemble #1
followed by several reduced discharge format ensembles.
When the ensemble is evenly divided by QR a full PD0 en-
semble is output.
Example:
PD = 3, QR = 4. Ensemble output order: FE, RD, RD, FE,
RD, RD, RD, FE, RD, RD, RD, FE, where FE = Full Ensem-
ble and RD is a reduced discharge output ensemble.
QU - Channel Geometry Units
Purpose:
Sets the units the QH and QW (distances, not epsilons) are
entered in.
Format: QU
n
Range:
n
= 0, 1 (0 = cm, 1 = mm)
Default: QU0
Description: Allows the ADFM to be set up using centimeters or millime-
ters for heights, widths, or radii.
ADFM Technical Manual (January 2000)
page C-35
Summary of Contents for ADFM Analog Output Module
Page 2: ......
Page 6: ...Table of Contents NOTES iv MGD Technologies Inc ...
Page 13: ...Table of Contents NOTES ADFM Technical Manual January 2000 xi ...
Page 14: ...Table of Contents xii MGD Technologies Inc ...
Page 54: ...Chapter 5 NOTES page 5 10 MGD Technologies Inc ...
Page 132: ...Appendix C NOTES page C 52 MGD Technologies Inc ...
Page 152: ...Appendix E NOTES page E 14 MGD Technologies Inc ...
Page 154: ......
Page 155: ......
Page 156: ......