Chapter 2
Installation of the ADFM transducer assembly must also comply with the
following parameters to maintain the accuracy of the final installation:
•
Transducer must be installed in such a way that the vertical beam
(beam 5) is oriented normal with respect to the pipe’s or chan-
nel’s invert. Note that the vertical beam is normal to the trans-
ducer’s top surface.
•
Transducer must not be rotated about the vertical Z-axis. A sight-
ing device or similar should be used to minimize rotational mis-
alignment.
•
Transducer must be installed such that positive longitudinal flow
direction is from the “non-cable” side of the transducer to the
“cable” side. See Figure 2-38 below for further explanation of
the flow direction.
Positive
(Downstream)
direction of flow
+x
+v
+z out of page
TOP
Figure 2-3.
Transducer Orientation
2-5.3
Install Software and Configure Station File
Before the ADFM can be used in a specific application, it must be
programmed for that application. This is most easily done using the
WinADFM software. See
Chapter 3
and the
Software Manual
for further
details.
page 2-8
MGD Technologies, Inc.
Summary of Contents for ADFM Analog Output Module
Page 2: ......
Page 6: ...Table of Contents NOTES iv MGD Technologies Inc ...
Page 13: ...Table of Contents NOTES ADFM Technical Manual January 2000 xi ...
Page 14: ...Table of Contents xii MGD Technologies Inc ...
Page 54: ...Chapter 5 NOTES page 5 10 MGD Technologies Inc ...
Page 132: ...Appendix C NOTES page C 52 MGD Technologies Inc ...
Page 152: ...Appendix E NOTES page E 14 MGD Technologies Inc ...
Page 154: ......
Page 155: ......
Page 156: ......