
Maintenance
4-5
Changing Mains Supply Voltage
If you need to change the mains supply voltage selection from 230V to
115V or vice-versa, follow the procedure below.
CAUTION.
Disconnect the mains power cord before attempting to change
the voltage selection. Make sure you install a fuse appropriate for the
selected mains supply.
a. Disconnect the mains power cord before attempting this procedure.
b. In a non-corrosive atmosphere, remove the top of the unit.
c. Locate plug P4 on the ADFM motherboard. Use the following table
to determine the correct jumpers.
Voltage Selection Jumper
115 VAC
P4.2 to P4.5
P4.4 to P4.6
230 VAC
P4.5 to P4.6
d. Label the electronic housing with a label indicating the proper mains
supply voltage.
e. Install a fuse appropriate for the selected mains supply (see
Table 4-16, item 10 or 11
).
f. Re-connect
the
mains power cord.
g. Test the system (see
Chapter 5
).
ADFM Technical Manual (January 2000)
page 4-5
Summary of Contents for ADFM Analog Output Module
Page 2: ......
Page 6: ...Table of Contents NOTES iv MGD Technologies Inc ...
Page 13: ...Table of Contents NOTES ADFM Technical Manual January 2000 xi ...
Page 14: ...Table of Contents xii MGD Technologies Inc ...
Page 54: ...Chapter 5 NOTES page 5 10 MGD Technologies Inc ...
Page 132: ...Appendix C NOTES page C 52 MGD Technologies Inc ...
Page 152: ...Appendix E NOTES page E 14 MGD Technologies Inc ...
Page 154: ......
Page 155: ......
Page 156: ......