TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
7-7
Table 7-6: Manual Span Check Procedure - Z/S Valves
Step Number
Action
Comment
1.
Press CALS
The M9110AH enters the span calibrate mode and switches the
Z/S valves to admit NO span gas.
2.
Select LO or HI
range
If AutoRange is enabled you will be prompted for which range
to calibrate. Select the LO or HI range, then press ENTR
3.
Wait 10 min
Wait for reading to stabilize at span value.
4.
Press SPAN
If you change your mind after pressing SPAN, you can still
press EXIT here without spanning the instrument.
5.
Press ENTR
Pressing ENTR actually changes the calculation equations and
causes the instrument to read the expected NO and NO
x
span
concentrations.
6.
Press EXIT
The Zero/Span valves are switched back to admit gas through
the sample port and instrument returns to SAMPLE mode.
Immediately after calibration, data is not added to the DAS
averages.
7.3 Dynamic Zero/Span Calibration Using AutoCal
The M9110AH can be automatically calibrated periodically using the Zero/Span Valve Option
and features in the AUTOCAL menu. The Dynamic Zero/Span feature can be enabled when an
AutoCal Sequence is set up. Refer to Section 6.2 for information.
With the Dynamic AutoCal in operation, the M9110AH is being continuously re-calibrated. We
recommend that the quality of the calibrations be carefully monitored to assure that the
instrument is performing correctly. Subtle fault conditions can sometimes be masked by
continuous re-calibration.
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...