TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
9-14
Table 9-4: Diagnostic Mode - Signal I/O (Continued)
No.
Signal
Control
Description
22
ST_FLOW_ALARM
YES
Status Bit - Flow alarm
Logic High = Sample/Ozone flow out of spec
Logic Low = Flows within spec
23
ST_TEMP_ALARM
YES
Status Bit - Temperature alarm
Logic High = Rxcell, Conv, Box temps out of spec
Logic Low = Temps within spec
24
ST_DIAG_MODE
YES
Status Bit - In Diagnostic mode
Logic High = M9110AH in Diagnostic mode
Logic Low = Not in Diag mode
25
ST_POWER_OK
YES
Status Bit - Power OK
Logic High = Instrument power is on
Logic Low = Instrument power is off
26
ST_SYSTEM_OK
YES
Status Bit - System OK
Logic High = No instrument warnings present
Logic Low = 1 or more alarms present
27
ST_HIGH_RANGE
YES
Status Bit - Autorange High Range
Logic High = M9110AH in high range of autorange
mode
Logic Low = M9110AH in low range of autorange
mode
28
PMT_SIGNAL
NO
Current PMT voltage. Same as PMT voltage in TEST
menu. Bi-polar, typically in 0-5000 mV range. A
constant value of 5000 mV indicates offscale.
29
SAMPLE_PRES
NO
Sample pressure in mV. Typical sea level value = 4300
mV for 29.9" Hg-A.
30
RCELL_TEMP
NO
Reaction Cell temperature. Typically 3500 mV for
50 C.
31
BOX_TEMP
NO
Box Temperature. Typically 1800 mV for 25 C
32
BLOCK_TEMP
NO
Temperature of the sample flow control orifice.
Typically 3500 mV for 50 C.
(table continued)
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...