TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
9-10
Table 9-2: Front Panel Warning Messages (Continued)
Message
Description
BLOCK TEMP WARNING
The sample flow orifice is located in a separate temperature
controlled block near the rear of the instrument. This message is
most often present during warm-up. It can also occur if heater is
not plugged in, or if thermistor has fallen out of block.
DCPS WARNING
DCPS is out of specification 2500 mV 200 mV. Check each
DCPS voltage per Section 9.3.5.
OZONE GEN OFF
Ozone generator is turned off. See Table 9-10 for conditions.
CANNOT DYN ZERO
Dynamic zero cal. out of spec. The reading of the PMT was too
high for the ZERO button to appear. Make sure the instrument is
receiving zero gas. Check for dirty reaction cell. Do the factory
calibration procedure located in Section 9.1.6.
CANNOT DYN SPAN
Dynamic span cal. out of spec. The reading of the PMT was too
high or low for the SPAN button to appear. Make sure the
instrument is receiving correct concentration span gas. Make
sure the expected span concentration is entered. Check for dirty
reaction cell. Do the factory calibration procedure located in
Section 9.1.6.
A/D NOT INSTALLED
V/F board has failed. The V/F board did not respond to
commands from the CPU. This probably means 1. Board not
seated in socket, 2. defective board, 3. defective back plane
connector.
9.1.3 Fault Diagnosis using DIAGNOSTIC Mode
Diagnostic mode can be looked at as a tool kit of diagnostics to help troubleshoot the
instrument.
To enter DIAG mode press:
SETUP-MORE-DIAG
The diagnostic modes are summarized in Table 9-3. To access these functions, after you have
pressed SETUP-MORE-DIAG, then press NEXT, PREV to select the desired mode, then press
ENTR to select the desired mode. This section is a detailed description of the test and
suggestions for its use.
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...