TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
9-13
Table 9-4: Diagnostic Mode - Signal I/O (Continued)
No.
Signal
Control
Description
11
OPTIC_TEST
YES
Turns on optic test bit in preamp. Should be used for
troubleshooting Preamp logic lines. We recommend
you use the OT button in the DIAG menu to operate
optic test.
12
BROWNOUT_RESET
YES
Brownout reset works in conjunction with
DSP_BROWNOUT. When DSP_BROWNOUT is set
the CPU sends a signal to reset the display and clear the
DSP_BROWNOUT.
13
CONV_HEATER
NO
Shows the status of the converter heater. This has the
same function as the LED in the PSM.
14
O3GEN_STATUS
YES
Switches power to the ozone generator. Can be used to
force the ozone generator to turn on.
15
ST_LOW_SPAN_CAL
YES
Status Bit - state of the low concentration span valve
Logic high = valve open to admin span gas
Logic low = valve closed
16
AUTOZERO_VALVE
YES
Switches the AutoZero valve. Use this bit to test the
valve function.
17
ST_RCELL_PRESS
YES
Status Bit - Reaction Cell Pressure alarm
Logic high = pressure out of acceptable range
Logic low = pressure inside acceptable range
18
PREAMP_RANGE_HI
YES
Switches the preamp hardware range. Standard ranges
are 500 and 5,000 ppm. Logic high = 5,000 ppm; logic
low = 500 ppm. M9110AH will reset range to correct
value based on user set range value.
19
ST_CONC_ALARM
YES
20
ST_ZERO_CAL
YES
Status Bit - Zero Calibration mode
Logic high = M9110AH in Zero cal mode
Logic low = Not in Zero cal mode
21
ST_SPAN_CAL
YES
Status Bit - Span Calibration mode
Logic high = M9110AH in Span cal mode
Logic low = Not in Span cal mode
(table continued)
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...