TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
6-3
Table 6-2: Setup Automatic Zero/Span Checking or Calibration
Step
Action
Comment
1.
Press SETUP-ACAL
This button sequence will cause the AUTOCAL menu to be displayed.
2.
Press PREV or
NEXT
This selects a SEQUENCE. Up to 3 Sequences can be defined. A
Sequence is a definition of zero/span calibrations along with timing
information.
3.
Select MODE, then
PREV or NEXT
MODE allows you to select the order of zero, low-span or high-span
that will be performed, MODES are listed in Table 6-3. Use PREV -
NEXT to scroll the list of choices, then press ENTR to select. To
disable a sequence, select DISABLE from the menu.
4.
Press SET
This menu sets up several timers and selects calibration or calibration
check.
1. TIMER ENABLE – If set to ON, the defined sequence will execute.
If set to OFF, the sequence will remain in memory, but will not
execute.
2. START DATE - calendar date to start sequence dd-mmm-yy, day-
month-year format.
3. START TIME - time of day sequence will start. hh-mm, hours-
minutes in 24 hr format.
4. DELTA DAYS and DELTA HOURS:MIN is the time between each
activation of the sequence. By setting the Delta Hours variable to be
other than an exact number of days, the sequence can move ahead or
be delayed each time it is run. For example, by setting DELTA
DAYS = 1 and DELTA HOURS = 1:00, the sequence will execute
1 hour later each day. To set it to move back ½ hour each day,
DELTA DAYS = 0, DELTA HOURS = 23:30.
5. DURATION -Amount of time for each zero/span point.
6. CALIBRATE – If set to ON, will calibrate(adjust slope and offset).
If OFF will check, but not change calibration.
7. 7. RANGE-TO-CAL Selects the range to calibrate LO or HI.
Applicable to AutoRange and RemoteRange modes.
5.
Press EXIT
The M9110AH now exits the AUTOCAL SETUP menu. For each
enabled sequence, the M9110AH will now AUTOCAL at the defined
times.
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...