9.1.3.1 Signal I/O Diagnostics
Table 9-4: Diagnostic Mode - Signal I/O
No.
Signal
Control
Description
0
DSP_BROWNOUT
NO
Display brownout is used to keep the display from
getting corrupted during low line voltage conditions.
Circuitry on the Status/Temp board senses low line
voltage and sets this bit. The CPU reads this and
generates the BROWNOUT_RST signal described
below.
1
EXT_ZERO_CAL
NO
Shows state of status input bit to cause the M9110AH
to enter Zero Calibration mode. Use to check external
contact closure circuitry.
2
EXT_SPAN_CAL
NO
Shows state of status input bit to cause the M9110AH
to enter the Span Calibration mode. Use to check
external contact closure circuitry.
3
EXT_LOW_SPAN
NO
Shows state of status input bit to switch the low span
concentration valve in the zero/span valve option
4
REMOTE_RANGE_HI
NO
Shows state of status input bit to switch the remote
range control from low to high range.
5
SPAN_VALVE
YES
Switches the Span valve. Use this bit to test the valve
function.
6
CAL_VALVE
YES
Switches the Cal valve. Use this bit to test the valve
function.
7
NOX_VALVE
YES
Switches the NO/NO
x
valve. Use this bit to test the
valve function.
8
RCELL_HEATER
YES
Shows the status of the reaction cell heater. This has
the same function as the LED in the PSM.
9
BLOCK_HEATER
YES
Shows the status of the sample flow orifice block
heater. This has the same function as the LED in the
PSM.
10
ELEC_TEST
YES
Turns on electric test bit in preamp. Should be used for
troubleshooting Preamp logic lines. We recommend
you use the ET button in the DIAG menu to operate
electric test.
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...