TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
2-7
7. When the instrument is warmed up, re-check the TEST functions against Table 2-1. All of
the readings should compare closely with those in the table. If they do not see Section 9.1.1.
The next task is to calibrate the analyzer. There are several ways to do a calibration, they are
summarized in Table 7-1. For a preliminary checkout we recommend calibration with span
gas coming in through the sample port. The procedure is:
Step 1 - Set the range, then enter the expected NO
x
and NO span gas concentrations:
Step Number
Action
Comment
1.
Press
CAL-CONC-NOX
This key sequence causes the M9110AH to prompt for the
expected span concentration. Enter the span value by pressing
the key under each digit until the expected value is set.
2.
Press ENTR
ENTR stores the expected NO
x
span value. The internal
formulas are adjusted to compute this number when span gas
concentration is input into the instrument.
3.
Press
CAL-CONC-NO
In the same CAL-CONC sub menu press the NO button and
enter the expected NO span value. As before this value will be
used in the internal formulas to compute the NO concentration
value.
4.
Press ENTR
ENTR stores the expected NO span value. The internal
formulas are adjusted to compute this number when span gas
concentration is input into the instrument.
5.
Press EXIT
Returns instrument to SAMPLE mode.
6.
Press
SETUP-RNGE-
MODE-SNGL
If necessary, you may want to change ranges. Normally the
instrument is shipped in single range mode set at 100 ppm. We
recommend doing the initial checkout on the 100 ppm range.
7.
Press
SETUP-RNGE-
SET
After SETUP-RNGE-SET, enter 100 and press ENTR. The
instrument will now be in the 100 ppm range.
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...