TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
9-49
PSM Diagnostic procedures
The Linear Power Supply board can be tested by checking the DCPS - TEST function on the
front panel. It should read 2500 mV 200 mV. If the value is outside this range, individual
output voltages can be tested on connector P3, see the schematic in the Appendix for pinouts.
The Switching Power Supply output can be tested by observing the red LED on the TEC
control board attached to the inside of the fan duct on the Sensor Module. If the LED is on, +15
VDC is being supplied to the cooler. The current draw through the cooler can be measured
across TP1 and TP2. The value of the current in amps is 10x the voltage measured, and is
typically 1.5 to 2.0 amps at room temperature.
The Switch Board can be tested by observing the diagnostic LEDS along the top edge of the
board. The following Table 9-9 describes the typical operation of each LED.
Table 9-9: Power Supply Module LED Operation
No.
Function
Description
1.
NO/NO
x
Valve
Should switch about every 6 sec. On = NO
x
mode Off =
NO mode.
4.
AutoZero Valve
Should switch every 2 minutes. On = Azero Off = Sample
mode
5.
Ozone Generator Power
Power to ozone generator is off for 30 minutes after
turning on an instrument that has been off for a long time.
If the instrument has been turned off for less than 60
minutes, the ozone generator is immediately turned on.
6.
Block Heater
Should cycle ON-OFF every 20 sec to 2 min. On
continuously until up to temp.
7.
Conv Heater
Should cycle ON-OFF every 20 sec to 2 min. On
continuously until up to temp.
8.
Reaction Cell Heater
Should cycle ON-OFF every 20 sec to 2 min. On
continuously until up to temp.
9.3.6 Ozone Generator
The ozone generator subsystem consists of a permeation drier, flowmeter, transformer -
generator assembly, and driver board. The location of the components is illustrated in Figure 9-
7. Ozone is generated by drying ambient air, then passing the air between two electrodes that
have a large oscillating electric field generated by a high voltage transformer.
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...