TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
6-4
Table 6-3: Action of MODE Field in AutoCal
MODE
CALIBRATE = ON
CALIBRATE = OFF
ZERO
Zero calibrates LO or HI (depending on
RANGE-TO-CAL), SINGLE or IND ranges.
Zero checks LO or HI (depending on
RANGE-TO-CAL), SINGLE or IND range.
ZERO-LO
Zero and low span calibrates LO, SINGLE or
IND ranges.
Zero and low span checks LO, SINGLE or
IND ranges.
ZERO-HI
Zero cals & span cals HI, SINGLE or IND
range with HI span gas
Zero checks & span checks on HI, SINGLE
or IND range with HI span gas
ZERO-LO-HI
Zero cals, span cals LO and HI range.
LO span gas for LO range, HI span gas
for HI range.
Zero cals & span cals in SINGLE or
IND range using HI span gas. LO span is
cal check only.
Zero checks, span checks both ranges.
LO span gas for LO range, HI span gas
for HI range. Does 3 point check in
SINGLE or IND range modes.
LO-HI
Span cals LO range with LO span gas,
Span cals HI range with HI span gas in
AutoRange mode. Span cals SINGLE or
IND range with HI span gas. LO span is
cal check only.
Span checks LO range with LO span
gas, Span checks HI range with HI span
gas. Does 2 point check in SINGLE or
IND range mode.
LO
Span cals LO range with LO span gas.
Span cals SINGLE or IND range with
LO span gas.
Span checks LO range with LO span
gas. Span checks SINGLE or IND range
with LO span gas.
HI
Span cals on HI range with HI span gas.
Span cals SINGLE or IND range with
HI span gas.
Span checks on HI range with HI span
gas. Span checks SINGLE or IND range
with HI span gas.
6.3 Isolated 4-20 mA Current Loop Option
The M9110AH comes equipped with 4-20mA current loop. The electrical ground of this output
is part of the instrument ground. In certain applications the instrument ground must be isolated
from the external ground. For applications requiring electrical isolation, we offer optional
isolated current loop outputs. With the exception of electrical isolation, the setup and usage is
identical to the standard equipment current loop output. The calibration procedure is located in
the Troubleshooting Section 9.
Summary of Contents for 9110AH
Page 2: ......
Page 11: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F x INTENTIONALLY BLANK ...
Page 13: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 1 2 INTENTIONALLY BLANK ...
Page 17: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 4 Figure 2 2 Rear Panel ...
Page 23: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 9 Figure 2 4 Front Panel ...
Page 24: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 11 ...
Page 28: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 2 15 INTENTIONALLY BLANK ...
Page 32: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 3 4 INTENTIONALLY BLANK ...
Page 35: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 4 3 Figure 4 1 Block Diagram ...
Page 45: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 5 3 Figure 5 2 Setup Menu Tree ...
Page 88: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 6 6 INTENTIONALLY BLANK ...
Page 102: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 7 14 INTENTIONALLY BLANK ...
Page 181: ...TELEDYNE Model 9110AH NOX Analyzer Operator Manual 01620 Rev F 9 65 INTENTIONALLY BLANK ...